Miscellaneous

How do you test Rapicide test strips?

How do you test Rapicide test strips?

Wait exactly thirty (30) seconds and then compare the test strip color with the color chart located on the side of the test strip bottle. Test results showing PAA levels of 1000 ppm and higher are defined as Positive. RAPICIDE® is a registered trademarks of Medivators Inc.

How long is Rapicide Pa Good For?

Please contact your representative for more information. 14 day reuse life so long as the minimum recommended concentration level, as read by the RAPICIDE PA and PA RTU Test Strips, reads above MRC of 1000ppm of PAA.

What is Rapicide made of?

RAPICIDE™ PA High-Level Disinfectant is a single-use, peracetic acid-based solution that is environmentally responsible with proven materials compatibility. It is designed for use in the ADVANTAGE PLUS™ Pass-Thru, ADVANTAGE PLUS™ or DSD EDGE™ Automated Endoscope Reprocessors.

How long is Rapicide good for after opening it?

Preparation and Usage: RAPICIDE OPA/28 is supplied ready-to-use and does not require activation. After opening, the solution remaining in the opened container may be stored for up to 75 days until used.

What is Rapicide PA?

Disinfects in five minutes at 86°F (30°C) Effectively kills TB, hepatitis viruses, Clostridium difficile, VRE, MRSA and CRE. Single-use, peracetic acid-based solution that is safe for the environment.

What is Rapicide used for?

INTENDED USE: RAPICIDE™ OPA/28 High-Level Disinfectant is intended to be used as a reusable liquid disinfectant solution for high-level disinfection of endoscopes and accessories in compatible automated endoscope reprocessing systems (AERs) and for manual disinfection of compatible medical and dental devices.

What is Rapicide Opa?

RAPICIDE™ OPA/28 is an ortho-Phthalaldehyde based high-level disinfectant that features a long reuse life. With the same contact time and temperature for manual disinfection as in an automated endoscope reprocessor, RAPICIDE OPA/28 provides fast turnaround time for endoscopes.

Is Rapicide a glutaraldehyde?

RAPICIDE™ High-Level Disinfectant and Sterilant is a fast-acting, non-staining disinfectant that requires no mixing or activation. This reusable glutaraldehyde disinfectant is designed for use on flexible endoscopes in automated endoscope reprocessors. Its fast contact time allows for quick endoscope turnaround time.

How do you dispose of Rapicide in PA?

Dispose of this product in accordance with all applicable Federal, State and Local regulations. If allowed by federal, state or local regulatory authority, flush spilled solution of RAPICIDE PA (Solutions A & B combined) to the drain using cold water. Run cold water for 15 minutes. Do not reuse the empty containers.

Does Rapicide have glutaraldehyde?

RAPICIDE High-Level Disinfectant and Sterilant, at a minimum recommended concentration (MRC) of 1.5% glutaraldehyde, has demonstrated efficacy in the presence of organic matter and microbiological burden during reuse.

How do you dispose of Rapicide?

What is MetriCide used for?

MetriCide OPA Plus is a high-level disinfectant for reprocessing heat-sensitive semi-critical medical devices for which sterilization is not suitable when used according to the instructions for use.

Who manufactures cidex Opa?

McKesson Medical-Surgical Inc.

How do you clean up a Rapicide spill?

If allowed by federal, state and/or the local regulatory authorities, flush neutralized Rapicide to the sewer with copious amounts of water. If a drain is not immediately available, collect the spill by using mops, towels or similar material to clean up the spill.

Does baking soda neutralize peracetic acid?

Bubbling of the bicarbonate with the material shows the neutralization occurring. If the tubs are not in a ventilated cabinet, leave the room. After allowing 5 minutes for the bicarbonate to neutralize the hydrogen peroxide/peracetic acid disinfectant, carefully pour the neutralized disinfectant down the drain.

Is MetriCide a high level disinfectant?

MetriCide 28 is a high-level disinfectant when used or reused, according to Directions for Use, up to 28 days at 25 °C, assuming the Minimum Effective Concentration (MEC) of glutaraldehyde, as measured by a chemical indicator, remains within acceptable parameters and other conditions of use are met, with an immersion …

What is intermediate level disinfection?

Intermediate-level disinfection refers to the use of an agent that kills a wider range of pathogens than a low-level disinfectant but does not kill bacterial spores. EPA-registered hospital disinfectants with a tuberculocidal claim are intermediate-level disinfectants.

How many test strips are in a pack of rapicide PA?

RNSML020118BX RAPICIDE PA Indicator Test Strip, 2 Bottles per Pack, 100 Test Strips in Each Bottle 2/BX Stock Allocated QTY: / Remaining QTY: / Allocation Reset Date: Specifications Latex Free Yes UNSPSC 42281806 Documents/SDS Category Id: – Content Id: Company About Us Careers Medline University Partner with Medline Patent Information

What is the life of the rapicide PA RTU RTU test strips?

14 day reuse life so long as the minimum recommended concentration level, as read by the RAPICIDE PA and PA RTU Test Strips, reads above MRC of 1000ppm of PAA. Requires only 1 rinse, reducing the total cycle time

What is rapicide™ PA for endoscopes?

It is designed for use on flexible endoscopes in automated endoscope reprocessors and manual applications. This solution is only available for International use outside of the USA, Canada and RAPICIDE™ PA Ready-to-Use High-Level Disinfectant and Sterilant | Cantel

What is the product bulletin for DSD edge rapicide PA?

Product Bulletin: DSD EDGE Automated Endoscope Reprocessor single rinse cycles with RAPICIDE PA High-Level Disinfection 50098-737 Rev A Product Bulletin: CRE-ERCP Statement 50098-711 Rev A Product Bulletin: RAPICIDEHigh-Level Disinfectant Efficacy against CRE 50098-712 Rev A Product Bulletin: Elevator/Raiser Wire Bridge Position During Reprocessing